Accuracy of MRI volume measurements of breast lesions: comparison between automated, semiautomated and manual assessment

Vertical Tabs

Eur Radiol
2009 May
19
5
1097-107
10.1007/s00330-008-1243-z
Epub Date: 
Thursday, January 22, 2009
Journal Articles
PubMed ID: 
19159935

The aim of this study was to investigate the efficacy of a dedicated software tool for automated and semiautomated volume measurement in contrast-enhanced (CE) magnetic resonance mammography (MRM). Ninety-six breast lesions with histopathological workup (27 benign, 69 malignant) were re-evaluated by different volume measurement techniques. Volumes of all lesions were extracted automatically (AVM) and semiautomatically (SAVM) from CE 3D MRM and compared with manual 3D contour segmentation (manual volume measurement, MVM, reference measurement technique) and volume estimates based on maximum diameter measurement (MDM). Compared with MVM as reference method MDM, AVM and SAVM underestimated lesion volumes by 63.8%, 30.9% and 21.5%, respectively, with significantly different accuracy for benign (102.4%, 18.4% and 11.4%) and malignant (54.9%, 33.0% and 23.1%) lesions (p