Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty

Vertical Tabs

Magn Reson Med
2012 May
67
5
1210-24
10.1002/mrm.23097
Epub Date: 
Friday, August 19, 2011
Journal Articles
PubMed ID: 
21858868

Simultaneous multislice Echo Planar Imaging (EPI) acquisition using parallel imaging can decrease the acquisition time for diffusion imaging and allow full-brain, high-resolution functional MRI (fMRI) acquisitions at a reduced repetition time (TR). However, the unaliasing of simultaneously acquired, closely spaced slices can be difficult, leading to a high g-factor penalty. We introduce a method to create interslice image shifts in the phase encoding direction to increase the distance between aliasing pixels. The shift between the slices is induced using sign- and amplitude-modulated slice-select gradient blips simultaneous with the EPI phase encoding blips. This achieves the desired shifts but avoids an undesired "tilted voxel" blurring artifact associated with previous methods. We validate the method in 3× slice-accelerated spin-echo and gradient-echo EPI at 3 T and 7 T using 32-channel radio frequency (RF) coil brain arrays. The Monte-Carlo simulated average g-factor penalty of the 3-fold slice-accelerated acquisition with interslice shifts is