Error message

  • Deprecated function: Creation of dynamic property MergeQuery::$condition is deprecated in MergeQuery->__construct() (line 1357 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).

Eur J Neurol. 2009 Apr;16(4):461-7 doi: 10.1111/j.1468-1331.2008.02398.x.

Changes in cerebral blood flow after acetazolamide: an experimental study comparing near-infrared spectroscopy and SPECT

Schytz HW, Wienecke T, Jensen LT, Selb J, Boas DA, Ashina M.

Abstract

BACKGROUND AND PURPOSE: It is important to find a reliable and bedside method, which can estimate the cerebral blood flow (CBF) of patients in clinical settings. Estimation of CBF by calculating a blood flow index (BFI) using continuous wave near-infrared spectroscopy (CW-NIRS) and indocyanine green (ICG) as an i.v. tracer has been proposed to be a feasible and promising method. To validate if the BFI method can detect relative changes in CBF we compared data with the established method (133)Xenon single photon emission computer tomography ((133)Xe-SPECT).
METHODS: Ten healthy subjects were investigated before and after a bolus of acetazolamide. NIRS data were obtained using a multi source detector separation configuration in order to assess a corrected BFI (BFI(corr)) value, which attempts to eliminate contamination of skin blood flow.
RESULTS: Data obtained showed no significant correlation between CBF changes measured by (133)Xe-SPECT and BFI(corr) (0.133, P = 0.732). After acetazolamide, a 49% increase in CBF was detected using the (133)Xe-SPECT method, whereas no changes in any ICG variables were observed after acetazolamide.
CONCLUSION: The study shows that it is not possible to obtain reliable BFI data, which reflect changes in CBF after acetazolamide infusion, using the CW-NIRS and ICG method.

PMID: 19236469