J Cardiovasc Comput Tomogr. 2012 May-Jun;6(3):172-183.e2 doi: 10.1016/j.jcct.2012.04.003. 2012 Apr 07.

Coronary computed tomography angiography during arrhythmia: Radiation dose reduction with prospectively ECG-triggered axial and retrospectively ECG-gated helical 128-slice dual-source CT

Lee AM, Engel LC, Shah B, Liew G, Sidhu MS, Kalra M, Abbara S, Brady TJ, Hoffmann U, Ghoshhajra BB.

Abstract

BACKGROUND: Arrhythmia during coronary computed tomography angiography (coronary CTA) acquisition increases the risk of nondiagnostic segments and high radiation exposure. An advanced arrhythmia rejection algorithm for prospectively electrocardiogram (ECG)-triggered axial scans using dual-source CT (DSCT) examinations has recently been reported.
OBJECTIVE: We compared image quality and effective dose at DSCT examinations using prospectively ECG-triggered axial scanning with advanced arrhythmia rejection software (PT-AAR) versus retrospectively ECG-gated helical scanning with tube-current modulation (RG-TCM) during arrhythmia.
METHODS: This was a retrospective case-control study of 90 patients (43 PT-AAR, 47 RG-TCM) with arrhythmia (defined as heart rate variability [HRV] > 10 beats/min during data acquisition) referred for physician-supervised coronary CTA between April 2010 and September 2011. A subset of 22 cases matched for body mass index, HR, HRV, and other scan parameters was identified. Subjective image quality (4-point scale) and effective dose (dose length product method) were compared.
RESULTS: PT-AAR was associated with lower effective dose than RG-TCM (4.1 vs 12.6 mSv entire cohort and 4.3 vs 9.1 mSv matched controls; both P 28 beats/min (P CONCLUSIONS: In patients with variable heart rates, prospectively ECG-triggered axial DSCT with arrhythmia rejection algorithm is feasible and can decrease radiation exposure by ∼50% versus retrospectively ECG-gated helical DSCT, with preserved image quality.

PMID: 22542280