Radiology. 2008 Oct;249(1):301-6 doi: 10.1148/radiol.2483072003. 2008 Aug 18.

Calcified plaque: measurement of area at thin-section flat-panel CT and 64-section multidetector CT and comparison with histopathologic findings

Sarwar A, Rieber J, Mooyaart EA, Seneviratne SK, Houser SL, Bamberg F, Raffel OC, Gupta R, Kalra MK, Pien H, Lee H, Brady TJ, Hoffmann U.

Abstract

The purpose of this study was to assess the blooming artifacts in ex vivo coronary arteries at multidetector computed tomography (CT) and flat-panel-volume CT by comparing measured areas of calcified plaque with respect to the reference standard of histopathologic findings. Three ex vivo hearts were scanned with multidetector CT and flat-panel-volume CT after institutional review board approval. The area of calcified plaque was measured at histopathologic examination, multidetector CT, and flat-panel-volume CT. The plaque area was overestimated at multidetector CT by 400% (4.61/1.15) on average, and the predicted difference between the measurements was significant (3.46 mm(2), P = .018). The average overestimation of plaque area at flat-panel-volume CT was twofold (214% [2.18/1.02]), and the predicted difference was smaller (1.16 mm(2), P = .08). The extent of the blooming artifact in visualizing calcified coronary plaque is reduced by using flat-panel-volume CT.

PMID: 18710960