Error message

  • Deprecated function: Creation of dynamic property MergeQuery::$condition is deprecated in MergeQuery->__construct() (line 1357 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).

Eur J Radiol. 2006 Mar;57(3):380-3 doi: 10.1016/j.ejrad.2005.12.024. 2006 Jan 25.

Improved vessel morphology measurements in contrast-enhanced multi-detector computed tomography coronary angiography with non-linear post-processing

Ferencik M, Lisauskas JB, Cury RC, Hoffmann U, Abbara S, Achenbach S, Karl WC, Brady TJ, Chan RC.

Abstract

Multi-detector computed tomography (MDCT) permits detection of coronary plaque. However, noise and blurring impair accuracy and precision of plaque measurements. The aim of the study was to evaluate MDCT post-processing based on non-linear image deblurring and edge-preserving noise suppression for measurements of plaque size. Contrast-enhanced MDCT coronary angiography was performed in four subjects (mean age 55 +/- 5 years, mean heart rate 54 +/- 5 bpm) using a 16-slice scanner (Siemens Sensation 16, collimation 16 x 0.75 mm, gantry rotation 420 ms, tube voltage 120 kV, tube current 550 mAs, 80 mL of contrast). Intravascular ultrasound (IVUS; 40 MHz probe) was performed in one vessel in each patient and served as a reference standard. MDCT vessel cross-sectional images (1 mm thickness) were created perpendicular to centerline and aligned with corresponding IVUS images. MDCT images were processed using a deblurring and edge-preserving noise suppression algorithm. Then, three independent blinded observers segmented lumen and outer vessel boundaries in each modality to obtain vessel cross-sectional area and wall area in the unprocessed MDCT cross-sections, post-processed MDCT cross-sections and corresponding IVUS. The wall area measurement difference for unprocessed and post-processed MDCT images relative to IVUS was 0.4 +/- 3.8 mm2 and -0.2 +/- 2.2 mm2 (p

PMID: 16442768
| MGH/HST Martinos Center for Biomedical Imaging

Error

The website encountered an unexpected error. Please try again later.