Error message

  • Deprecated function: Creation of dynamic property MergeQuery::$condition is deprecated in MergeQuery->__construct() (line 1357 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).
  • Deprecated function: Creation of dynamic property DatabaseCondition::$stringVersion is deprecated in DatabaseCondition->compile() (line 1887 of /webdata/web/html/includes/database/query.inc).

J Magn Reson Imaging. 2013 Feb;37(2):457-70 doi: 10.1002/jmri.23823. 2012 Sep 25.

Single-breath xenon polarization transfer contrast (SB-XTC): implementation and initial results in healthy humans

Muradyan I, Butler JP, Dabaghyan M, Hrovat M, Dregely I, Ruset I, Topulos GP, Frederick E, Hatabu H, Hersman WF, Patz S.

Abstract

PURPOSE: To implement and characterize a single-breath xenon transfer contrast (SB-XTC) method to assess the fractional diffusive gas transport F in the lung: to study the dependence of F and its uniformity as a function of lung volume; to estimate local alveolar surface area per unit gas volume S(A)/V(Gas) from multiple diffusion time measurements of F; to evaluate the reproducibility of the measurements and the necessity of B(1) correction in cases of centric and sequential encoding.
MATERIALS AND METHODS: In SB-XTC three or four gradient echo images separated by inversion/saturation pulses were collected during a breath-hold in eight healthy volunteers, allowing the mapping of F (thus S(A)/V(Gas)) and correction for other contributions such as T(1) relaxation, RF depletion and B(1) inhomogeneity from inherently registered data.
RESULTS: Regional values of F and its distribution were obtained; both the mean value and heterogeneity of F increased with the decrease of lung volume. Higher values of F in the bases of the lungs in supine position were observed at lower volumes in all volunteers. Local S(A)/V(Gas) (with a mean ± standard deviation of S(A)/V(Gas) = 89 ± 30 cm(-1)) was estimated in vivo near functional residual capacity. Calibration of SB-XTC on phantoms highlighted the necessity for B(1) corrections when k-space is traversed sequentially; with centric ordering B(1) distribution correction is dispensable.
CONCLUSION: The SB-XTC technique is implemented and validated for in vivo measurements of local S(A)/V(Gas).

PMID: 23011916
| MGH/HST Martinos Center for Biomedical Imaging

Error

The website encountered an unexpected error. Please try again later.