J Magn Reson. 2007 Aug;187(2):177-83 doi: 10.1016/j.jmr.2007.04.011. 2007 Apr 27.

Improved diffusion measurement in heterogeneous systems using the magic asymmetric gradient stimulated echo (MAGSTE) technique

Sun PZ.

Abstract

A magic asymmetric gradient stimulated echo (MAGSTE) sequence was recently proposed to improve molecular diffusion measurements in the presence of spatially varying background gradients. Its effectiveness has been demonstrated previously with simulated background gradients and in phantoms that contain bulk susceptibility differences. In this study, we investigated the MAGSTE technique in microscopically heterogeneous systems, and compared it with the conventional bipolar pulsed gradient stimulated echo (bPGSTE) sequence. We demonstrated that the MASGTE measurements, compared to the bPGSTE method, varied significantly less when the diffusion encoding/decoding interval (delta) was changed. In addition, the MAGSTE technique provided good characterization of the surface area-to-volume ratio for heterogeneous systems investigated in this study. In sum, this study showed that the MAGSTE technique provided diffusion measurements superior to those of the bPGSTE sequence, especially in the presence of severe heterogeneous background gradients.

PMID: 17500019